66 Chapter 6: Sample Experiments
Calorimetry Sciences Corp.
CSC 5300 N-ITC III 67
User’s Manual
the tip of the injection needle during equilibration, which results in fewer moles of titrant
being injected in the rst injection. If you want to discard this point in the t then position
the mouse pointer directly over the point and click the left button once. The point should
be replaced by a red triangle, indicating that the integrated area will not be used in the t.
Next, click the Fit Selected Model button. The values for K, ∆H and n will be displayed
in a table located directly above a graph showing both the data and t. If you would like
statistical analysis of the t click the Statistical Analysis button.
Figure 6-6: Modeling the RNase A experiment.
Statistical analysis
Fit selected
model